Single-Sample Micro-Osmometer
1. Single-sample osmometer
2. Direct-draw loading
3. Secure and efficient data management
4. Supports 21 CFR part 11 & EU Annex 11 compliance
5. Meets Pharmacopeia osmolality testing guidelines
|
Sample type |
Aqueous-based solution |
|
Sample volume |
20 ± 1 μL |
|
Test time |
90 seconds |
|
Sample Capacity |
Single sample |
|
Unit |
mOsm/kg H2O |
|
Resolution |
1 mOsm/kg H2O |
|
Range |
0-2000 mOsm/kg H2O |
|
Accuracy2 |
0-400 mOsm/kg H2O: mean value ± 2 mOsm/kg H2O from nominal value (1 SD) >400- to <1500 mOsm/kg H2O:≤ 0.5% from nominal value (1 SD) >1500- to 2000 mOsm/kg H2O:≤ 1% from nominal value (1 SD) |
|
Within-run Repeatability |
0-400 mOsm/kg H2O: standard deviation ≤ 2 mOsm/kg H2O >400 to <1500 mOsm/kg H2O: Coefficient of Variation ≤ 0.5% >1500 to <2000 mOsm/kg H2O: Coefficient of Variation ≤ 1% |
|
Temperature effects3 |
< 1 mOsm/kg H2O per 5°C (9°F) ambient temperature change |
|
Calibration |
0*, 50, 850, and 2000* mOsm/kg H2O (* optional) |
|
Communication |
USB 2.0 Type A Port (2), USB 2.0 Type A Port (1), Ethernet 10/100, RJ45 connector port (1) dot matrix printer (optional) |
|
Supported Languages |
Simple Chinese, Czech, Danish, Greek, Portuguese, English, French, German, Italian, Japanese, Korean, Russian, Spanish, Turkish |
|
Storage temperature |
-20°C to +45°C (-40°F to +113°F) |
|
Electrical voltage |
100-240 VAC (50/60 Hz) |
|
Power consumption |
60W |
|
Dimensions (D x W x H) |
38 cm x 35 cm x 29 cm (15” x 14” x 11.5”) |
|
Net weight |
6.0 kg (13.3 lbs.) |
1.Subject to change
2.Accuracy and precision (within run) specifications apply to Advanced Instruments standardsand reference solutions.
Performance at Reference Conditions: 20°C to 25°C (68°F to 77°F);40 to 60% relative humidity
3.Operating Conditions: Temperature 18°C to 35°C (64°F to 95°F); 5 to 80% relative humidity(non-condensing)
4.Dimensions when Micro-Sample Test Kit is on the instrument