• Fiske® 210 Micro-Sample Osmometer

Fiske® 210 Micro-Sample Osmometer

Model : 210

1. Alphanumeric Display

2. Automatic Calibration

3. Easy push-button operation

4. smallest sample size

5. printer, RS232C and barcode scanner ports

6. Solid-state cooling


Sample volume

20

Test time

90 seconds

Sample Capacity

Single sample

Unit

mOsm/kg H2O

Resolution

1 mOsm/kg H2O

Range

0-2000 mOsm/kg H2O

Communications

DTE RS-232 serial port, parallel printer port, and optional barcode scanner

Linearity

Less Than ±1% from a straight line between 0 and 2000 mOsm/kg H2O

Repeatability

±2 mOsm/kg H2O (1 S.D.) between 0 to 400 mOsm/kg H2O

±0.5 %(1 S.D) between 400 to 2000 mOsm/kg H2O

Drift

Less Than 1 mOsm/kg H2O per month

Temperature Effects

Less than 1 mOsm/kg H2O per 5°C (9°F) ambient temperature change

Temperature

18 to 35°C (64 to 95°F)

Humidity

5 to 80% relative humidity, (non-condensing)

Storage temperature

-40°C to +45°C (-40°F to +113°F)

Electrical voltage

100-240 VAC (50/60 Hz)

Power consumption

60W

Dimensions

(D x W x H)

9.5″ x 12″ x 12″ 

(2.1 cm x 30.5 cm x 30.5 cm)

Net weight

6.2 kg (13.6 lbs.)